UNISCAN 3200 – Universal Input Scanner / Data Logger
Short description: A rugged microcontroller-based universal input scanner and data logger with up to 32 channels, real-time/batchwise logging, MODBUS RTU connectivity, and parallel printer support. Ideal for pharma and industrial automation applications.
No serial information available.
Key Features
- Compact, rugged microcontroller-based system.
- Accepts universal inputs: RTD Pt100, 0–20 mA, 4–20 mA, thermocouples (J, K, R, S, T, B, E), 0–1 V.
- Up to 32 channels (expandable in steps of 8).
- 16-point linearisation for RTD, mA, mV inputs.
- Alarm indication via front panel LEDs with common relay output.
- Real-time or batchwise data logging with manual start/stop and programmable batch number.
- Memory capacity:
- 1700 readings/channel (32-channel model)
- 5400 readings/channel (8-channel model)
- Real-time data logging capacity up to 50,000 readings.
- Data logging interval: user-settable (1–240 sec/min).
- Memory with 48-hour battery backup; memory full indication on LCD.
- RS232/RS485/RS422 with MODBUS RTU protocol for SCADA.
- Parallel port for Centronics-compatible dot-matrix printers.
- Batchwise printing and Excel-format PC transfer.
- Tag names, engineering units, Hi/Lo alarm limits per channel.
- Password protection for programming.
- Channel skip facility for display & logging.
- Ideal for pharmaceutical and process automation applications.
Technical Specifications
- Inputs: RTD Pt100 (3-wire), 0–20 mA, 4–20 mA, Thermocouples (J, K, R, S, T, B, E), 0–1 V (programmable per channel)
- Channels: Up to 32 channels in steps of 8
- Display: 20×2 backlit LCD
- Keyboard: 16-key tactile keypad
- Resolution: RTD/T-type: 0.1°C; mA: 0.01; Volt: 0.001; other TCs: 1°C
- Scan Rate: 1000 ms per channel
- Logging Interval: Programmable 1–240 sec/min
- Accuracy: ±0.5% of FSR ±1 bit ADC
- Memory: 1700 readings/channel (32-ch) or 5400 readings/channel (8-ch) with real-time stamping
- Special Features: 16-point linearisation, tag names, engineering units, Hi/Lo limits per channel, channel skip
- Alarms: Common relay output, individual LED alarms per channel
- LED Indications: Alarm, Serial Tx/Rx, Memory full
- Printing: Parallel printer port (DOT matrix), auto/manual printing, interval 1–240 min
- Communication: RS232/RS485/RS422 with MODBUS RTU or ASCII protocol
- Mounting: Panel mount; Bezel: 180 × 305 mm; Depth: 300 mm; Cut-out: 180 × 262 mm
- Connections: Male/Female connectors on rear panel for each channel
Support Videos
Installation Guide
Troubleshooting
Downloads
Need More Help?
Call us at +91-20-24479210 or email scan@vsnl.com