UNISCAN 3200

UNISCAN 3200 – Universal Input Scanner / Data Logger

Short description: A rugged microcontroller-based universal input scanner and data logger with up to 32 channels, real-time/batchwise logging, MODBUS RTU connectivity, and parallel printer support. Ideal for pharma and industrial automation applications.

No serial information available.


UNISCAN 3200 Universal Input Scanner / Data Logger
UNISCAN 3200 – Universal Input Scanner / Data Logger

Key Features

  • Compact, rugged microcontroller-based system.
  • Accepts universal inputs: RTD Pt100, 0–20 mA, 4–20 mA, thermocouples (J, K, R, S, T, B, E), 0–1 V.
  • Up to 32 channels (expandable in steps of 8).
  • 16-point linearisation for RTD, mA, mV inputs.
  • Alarm indication via front panel LEDs with common relay output.
  • Real-time or batchwise data logging with manual start/stop and programmable batch number.
  • Memory capacity:
    • 1700 readings/channel (32-channel model)
    • 5400 readings/channel (8-channel model)
  • Real-time data logging capacity up to 50,000 readings.
  • Data logging interval: user-settable (1–240 sec/min).
  • Memory with 48-hour battery backup; memory full indication on LCD.
  • RS232/RS485/RS422 with MODBUS RTU protocol for SCADA.
  • Parallel port for Centronics-compatible dot-matrix printers.
  • Batchwise printing and Excel-format PC transfer.
  • Tag names, engineering units, Hi/Lo alarm limits per channel.
  • Password protection for programming.
  • Channel skip facility for display & logging.
  • Ideal for pharmaceutical and process automation applications.

Technical Specifications

  • Inputs: RTD Pt100 (3-wire), 0–20 mA, 4–20 mA, Thermocouples (J, K, R, S, T, B, E), 0–1 V (programmable per channel)
  • Channels: Up to 32 channels in steps of 8
  • Display: 20×2 backlit LCD
  • Keyboard: 16-key tactile keypad
  • Resolution: RTD/T-type: 0.1°C; mA: 0.01; Volt: 0.001; other TCs: 1°C
  • Scan Rate: 1000 ms per channel
  • Logging Interval: Programmable 1–240 sec/min
  • Accuracy: ±0.5% of FSR ±1 bit ADC
  • Memory: 1700 readings/channel (32-ch) or 5400 readings/channel (8-ch) with real-time stamping
  • Special Features: 16-point linearisation, tag names, engineering units, Hi/Lo limits per channel, channel skip
  • Alarms: Common relay output, individual LED alarms per channel
  • LED Indications: Alarm, Serial Tx/Rx, Memory full
  • Printing: Parallel printer port (DOT matrix), auto/manual printing, interval 1–240 min
  • Communication: RS232/RS485/RS422 with MODBUS RTU or ASCII protocol
  • Mounting: Panel mount; Bezel: 180 × 305 mm; Depth: 300 mm; Cut-out: 180 × 262 mm
  • Connections: Male/Female connectors on rear panel for each channel

Support Videos

Installation Guide

Troubleshooting

Downloads

Need More Help?

Call us at +91-20-24479210 or email scan@vsnl.com

🔙 Go back to Home Page